Presidium FAQs & Support

Help with your Presidium instrument

Find practical guidance on preparing gemstones for testing, checking your instrument and understanding its results.

Start with our Presidium Gem Tester II (PGT II) guide and frequently asked questions below. For advice on other Presidium instruments, contact H.S. Walsh.

View the PGT II product and handbook downloads

🎥 See the Product in Action

See the tool in action before reading the full specification.

Presidium Gem Tester II (PGT II)

Walsh product code: TD53

The PGT II measures a gemstone’s thermal conductivity—how readily it conducts heat. Its analogue dial displays possible gemstone ranges to help you investigate a stone’s identity alongside other gemmological observations and tests.

Understanding its limitations

  • It cannot distinguish diamond from moissanite.
  • It cannot distinguish natural gemstones from synthetic counterparts, including natural and laboratory-grown diamonds.
  • Some gemstone ranges overlap. A reading alone does not provide conclusive identification.

Prepare the stone carefully, follow the handbook for your version of the instrument and check the reference readings before relying on test results.

View PGT II product Read PGT II FAQs

PGT II quick help and FAQs

Select a question to see the answer.

Before testing

  1. Connect the probe with the unit switched off, then allow the handbook warm-up period.
  2. Check the CAL and Glass reference readings.
  3. Use clean, dry stones that have reached room temperature.
Do I need to adjust calibration at every startup?

Check the reference readings after warm-up. Adjust calibration only as directed in the matching handbook, such as after a probe replacement or incorrect reference readings.

How should I clean a stone?
  • Start with a clean, soft cloth containing no polishing compounds. Avoid silver-polishing cloths.
  • For suitable items, use lukewarm water and a very soft brush. Dry thoroughly before testing.
  • Unsure about the stone, treatment or setting? Ask for advice before wet cleaning, solvents, steam or ultrasonic cleaning.

Understanding results

Readings indicate possible identities. Overlapping ranges need further tests.

Does a diamond-range reading prove it is diamond?

No. Diamond and moissanite can give similar readings. Further testing is needed to separate them.

Can it separate natural and laboratory-grown stones?

No. The PGT II cannot establish natural or laboratory-grown origin, including for diamonds.

Troubleshooting

Why do readings vary?
  • Recheck power, reference readings, cleanliness and temperature.
  • Use the probe contact and pressure described in the handbook.
  • Allow very small stones to cool before repeating tests.
Why is the metal alert sounding?

The probe may be touching the setting or another metal surface. Reposition it onto the gemstone and repeat the test.

What if the reference checks still fail?

Follow the matching handbook's calibration instructions. If checks still fail or the probe appears damaged, contact Walsh before relying on further readings.

For demonstrations, troubleshooting or repair enquiries, contact the H.S. Walsh team. Quote PGT II / TD53.

Handbooks and documents

Download the PGT II user handbook in your preferred language. These PDFs are hosted by H.S. Walsh.

Other handbook languages

Chinese / 中文, French / Français, Italian / Italiano and Russian / Русский handbooks are available on request.

Request a handbook by email, stating PGT II / TD53 and your preferred language.

Check that the handbook matches your version of the instrument. Contact Walsh if you are unsure.

Demonstrations, troubleshooting and Warranty

Contact the H.S. Walsh team for help with your Presidium instrument or advice on arranging a Demonstration or Warranty assessment.

Please include:

  • The instrument model and product code.
  • A brief description of the help you need.
  • Your order reference, if available.

For PGT II faults, include the CAL and Glass check results. A clear photo or short video may help.

Contact your Walsh branch Email H.S. Walsh

Not sure which Presidium tester is right for you? Explore our range of diamond and gemstone testers and find the perfect instrument for your needs with our easy-to-use guide.

Find My Tester

Diamond Mate A (PDMT-A)

Walsh code: TD5610

A portable thermal conductivity tester for initial diamond and simulant checks, powered by 3 × AAA alkaline batteries.

A diamond indication needs further testing: moissanite can give the same result. This tester cannot establish natural or laboratory-grown origin or grade a diamond.

View Diamond Mate A

Read FAQs  |  Download handbooks

Diamond Mate demonstration

See the tester in use. Follow the PDMT-A handbook for your battery-powered model.

Diamond Mate A quick help and FAQs

Getting started

What powers the PDMT-A?

Use 3 × AAA alkaline batteries or a compatible Presidium AC adaptor, sold separately. The PDMT-C charging instructions do not apply to this model.

When is it ready, and does it need calibration?

Wait until the green power light stops blinking. The PDMT-A is factory calibrated: do not attempt user calibration. Contact Walsh if its performance concerns you.

How should I hold the tester and stone?

Keep fingers on the serrated metal panels on both sides. Hold the probe perpendicular to the stone’s table and press gently until the tip retracts. For loose stones, use a metal stone rest, held in your other hand. A rest is not supplied.

Understanding results

Does a green diamond result prove what the stone is?

No. Moissanite can also produce a diamond indication. The PDMT-A cannot separate natural from laboratory-grown diamonds or grade a diamond. Follow up with appropriate further tests.

What do red and intermittent beeping mean?

A red result indicates a simulant response, without identifying the material. Intermittent beeping indicates metal contact: reposition the probe onto the gemstone, clear of the setting.

Care and troubleshooting

How should I prepare an unknown stone?

Start with a soft, dry, lint-free cloth without polishing compounds. Avoid silver polishing cloths. Check suitability before using water, solvents, steam or ultrasonic cleaning. Test only when clean, dry and at room temperature.

Why are my readings inconsistent?

Check battery power, probe contact and cleanliness. Let small stones cool between repeat tests. Replace weak batteries before testing; if problems persist, ask Walsh for an assessment.

PDMT-A handbooks  |  Contact the Walsh team

Diamond Mate A handbooks

Use the QR code card supplied with your tester or choose a handbook below. These handbooks cover both PDMT-A and PDMT-C: follow the PDMT-A sections for your model.

Walsh-hosted multilingual handbook (PDF)

Presidium certificate of conformity (PDF)

For demonstrations, troubleshooting or warranty assessment, contact the Walsh team.

Adamas diamond and moissanite tester

Walsh code: TD5611

A handheld tester for colourless diamonds, moissanite and simulants, with replaceable magnetic probes. The regular probe is suitable from 0.02ct; a compatible micro probe extends testing to diamond melees from 0.01ct.

Diamond origin needs further testing. Adamas cannot establish natural or laboratory-grown origin or grade a diamond.

View Adamas

Read FAQs  |  Handbooks and probe guidance

Adamas quick help and FAQs

Power and probes

What powers Adamas, and does USB charge it?

Use 3 × AAA alkaline batteries, or the appropriate Presidium USB adaptor and cable. USB supplies operating power; it does not recharge the batteries.

What is the difference between the regular and micro probes?

Use the TZD5611A standard magnetic probe for diamond melees from 0.02ct, or the TZD5611B micro magnetic probe for melees as small as 0.01ct. Both fit this magnetic-tip Adamas.

What should I do after replacing a probe?

Switch the tester off before changing its magnetic tip. Fit the matching probe, then complete the handbook’s calibration procedure with the correct disk. Wait for the Ready light to stop blinking before testing. Ask Walsh if calibration fails.

Understanding results

Can Adamas separate natural and laboratory-grown diamonds?

No. A diamond indication does not establish origin. Further screening or gemmological testing is needed. Adamas also does not grade diamonds.

Will every simulant give a reading?

No. The handbook states that glass and cubic zirconia may give no reading. The simulant indication does not identify an individual gemstone species. Follow the handbook and investigate uncertain results.

Troubleshooting

Why does the metal alert appear?

The probe may be touching a claw or another metal surface. Reposition it onto the gemstone, clear of the setting.

Why are readings inconsistent?

Check power, probe fitting and calibration. Use a plain, soft, dry cloth without polishing compounds. Let stones reach room temperature and small stones cool between tests. Keep fingers on the grip and use the handbook’s probe contact and loose-stone support method. Contact Walsh if problems persist.

Adamas handbook information  |  Contact the Walsh team

Adamas handbooks – magnetic-tip version

These manufacturer handbooks cover the magnetic-tip Adamas. Choose your language below. Follow the calibration instructions after fitting a replacement magnetic probe.

Presidium certificate of conformity (PDF)

If your Adamas has an earlier twist-lock probe fitting, ask the Walsh team for the matching guide and replacement advice.

Presidium SAM demonstration

Watch Presidium’s demonstration, then follow the SAM handbook for setup and testing.

SAM diamond and moissanite tester

Walsh code: TD5613

A handheld tester for colourless diamonds, moissanite and simulants, including a wider range of low electrical conductivity moissanites. Its 0.45mm probe tests colourless diamonds as small as 0.01ct with correct contact.

Diamond origin needs further testing. SAM cannot establish natural or laboratory-grown origin or grade diamonds. It is not intended for coloured diamonds.

View SAM

Read FAQs  |  Download handbooks

SAM quick help and FAQs

Prepare clean, dry stones with a plain, soft cloth without polishing compounds. Allow the item to reach the recommended testing temperature of 23–27°C.

Getting started

How is SAM powered?

Use 3 × AAA alkaline batteries, or the appropriate Presidium USB adaptor and cable. The mains adaptor is an optional accessory sold separately.

When is it ready, and does it need calibration?

Allow approximately 10 seconds for warm-up and wait for the blue Ready light to stop blinking. SAM is factory calibrated; do not attempt user calibration.

Understanding results

Does a diamond result establish natural origin?

No. Natural and laboratory-grown diamonds can give a diamond indication. Further origin screening or gemmological testing is needed. SAM does not grade diamonds.

What do the lights mean?

Read the segment label as well as its colour.

SignalMeaning
Green Diamond + continuous beepDiamond indication
Yellow MoissaniteMoissanite indication
Red SimulantSimulant response; species not identified
Orange Metal + intermittent beepMetal contact
Red Low BattReplace batteries before testing
No resultGlass or cubic zirconia may give no reading; check setup and contact too

Testing and troubleshooting

Why do diamond and moissanite readings alternate?

Follow the handbook to test different points, clear of the setting. Let small stones cool between tests. If readings continue to alternate, refer the stone to a reputable gemmological laboratory.

How should I support loose stones?

Use the handbook’s metal stone-rest method: hold the rest in one hand and the tester in the other. Keep fingers on the tester’s metal grip and the probe perpendicular to the facet.

Can I replace the probe myself?

The handbook advises against removing the SAM tip. Contact Walsh for advice or assessment if the probe is damaged or results remain unreliable.

SAM handbooks  |  Contact the Walsh team

SAM handbooks and documents

Choose a manufacturer handbook below. Check it matches your SAM before following the operating instructions.

Presidium declaration of conformity (PDF)

For demonstrations, troubleshooting or warranty assessment, contact the Walsh team.

Presidium Gem Indicator video

Watch the video featured on Presidium’s PGI page. Follow the matching handbook for operating instructions.

Gem Indicator PGI

Walsh code: TD5612

A handheld thermal tester that indicates possible gemstone types. Select a colour yourself to help narrow the results. PGI covers up to 31 gemstone types and has a replaceable 0.45mm probe for stones from 0.02ct.

Further testing is needed to confirm identity. PGI cannot separate diamond from moissanite, establish natural or laboratory-grown origin, or grade diamonds.

View Gem Indicator PGI

Read FAQs  |  Handbooks and charts

PGI quick help and FAQs

Understanding results

Does a displayed gemstone name confirm its identity?

No. PGI suggests possible matches from thermal conductivity and your colour selection. Other stones can share those properties. It cannot separate diamond from moissanite or distinguish natural from laboratory-grown material. Confirm identity with supporting gemmological tests.

Does PGI measure the stone’s colour automatically?

No. You choose the colour from 12 options to narrow the possible matches. With the Color setting disabled, results use thermal conductivity only. The manufacturer’s colour chart is an approximate reference, not a colour-grading standard.

Testing and calibration

How can I get more consistent readings?

Use clean, dry stones at 23–27°C. Use a plain, soft cloth without polishing compounds. Keep fingers on the metal grip, hold the probe perpendicular to the facet and maintain contact until the progress bar finishes. Avoid metal settings. Allow small stones 5–10 seconds to cool between tests.

When does PGI need calibration?

PGI is factory calibrated and normally needs no routine adjustment. Follow the handbook when replacing a probe, investigating repeated incorrect results on known stones, or testing outside 23–27°C. For the latter, calibrate at the testing temperature and stay within 20–33°C.

The full procedure includes an initial automatic stage with the tip clear of all surfaces, then Disk 1 (grey) and Disk 2 (black). Follow every handbook step and confirm both disk stages succeed.

Can I replace the probe tip?

Yes. Presidium lists a PGI-specific replacement, GPROBE. Follow the matching PGI handbook for fitting and recalibration. Contact the Walsh team for availability and advice.

Why does SYS CHECK show Error?

The handbook lists poor probe connection, an uncalibrated replacement tip or damage as possible causes. Follow its instructions to reseat and recalibrate the probe. If the error persists, contact Walsh for assessment.

Power

How is PGI powered, and does USB charge it?

Use 3 × AAA alkaline batteries or the appropriate Presidium USB adaptor and cable. USB supplies operating power; it does not charge the tester or batteries. Wait for SYS CHECK to finish before testing. At one battery bar, replace the batteries; the handbook recommends recalibration at two bars.

PGI handbooks and charts  |  Contact the Walsh team

PGI handbooks and reference charts

Choose a manufacturer handbook below. Check it matches your PGI before following the operating instructions.

Reference documents

The charts help interpret possible matches; they do not confirm a gemstone’s identity.

For demonstrations, troubleshooting or warranty assessment, contact the Walsh team.